Journal of Korean Society of Ultrasound in Medicine 2012;31(4): 239-245.
Comparison of Diagnostic Performance of US Elastography and Conventional B-mode US in Differentiation of Breast Lesions.
Ji Young Kang, Jin Hwa Lee, Eun Kyung Kim, Suyoung Shin, Byoung Gwon Kim, Jin Han Cho, Dong Ho Ha, Byeong Ho Park, Sunseob Choi
1Department of Radiology, Dong-A University College of Medicine, Busan, Korea. jhrad@dau.ac.kr
2Department of Radiology, Yonsei University College of Medicine, Seoul, Korea.
3Department of Radiology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Korea.
4Department of Preventive Medicine, Dong-A University College of Medicine, Busan, Korea.
  Published online: December 31, 2012.
ABSTRACT
PURPOSE: The purpose of this study was to compare the diagnostic performance of ultrasound (US) elastography and conventional B-mode US for discrimination between benign and malignant breast lesions. MATERIALS AND METHODS: During a 13-month period, 277 women with 335 sonographically visible breast lesions who were scheduled to undergo biopsy were examined with US elastography. Elastographic findings were classified as benign or malignant based on the area ratio, with 1.00 as the threshold. Findings on conventional B-mode US were classified according to the BI-RADS category, as follows: lesions of BI-RADS categories 2 and 3 were considered benign, while those in categories 4 and 5 were considered malignant. Statistical analysis included sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and ROC curve analysis for comparison of the diagnostic performance of US elastography and conventional B-mode US. RESULTS: Of the 335 breast lesions, 85 (25.4%) showed malignancy on pathology. Findings on B-mode US showed malignancy in 264 (78.8%) and elastographic findings showed malignancy in 102 (30.4%). The sensitivity, specificity, PPV, NPV, and accuracy of B-mode US and elastography were 98.8%, 28.0%, 31.8%, 98.6%, and 79.4% and 69.4%, 81.2%, 57.8%, 88.8%, and 79.4%, respectively. Elastography showed significantly higher specificity and PPV and lower sensitivity and NPV, compared with B-mode US (p < 0.001). The area under the ROC curve (AUC value) was 0.761 for elastography, and 0.634 for B-mode US (p < 0.001). CONCLUSIONS: US elastography can improve specificity and PPV of B-mode US, but with significant sacrifice of sensitivity and NPV. Therefore, US elastography may complement B-mode US for differentiation of breast masses.
Keywords: Elastography; Breast US; Breast mass
TOOLS
Full text via DOI  Full text via DOI
METRICS
1,264
View
13
Download
We recommend
Editorial Office
A-304 Mapo Trapalace, 53 Mapo-daero, Mapo-gu, Seoul 04158, Korea
TEL : +82-2-763-5627   FAX : +82-2-763-6909   E-mail : office@ultrasound.or.kr
About |  Browse Articles |  Current Issue |  For Authors and Reviewers
Copyright © Korean Society of Ultrasound in Medicine.                 Developed in M2PI
Zoom in Close layer