Comparison of point and 2-dimensional shear wave elastography for the evaluation of liver fibrosis
Article information
Abstract
Purpose
This study aimed to assess the technical performance of ElastQ Imaging compared with ElastPQ and to investigate the correlation between liver stiffness (LS) values obtained using these two techniques.
Methods
This retrospective study included 249 patients who underwent LS measurements using both ElastPQ and ElastQ Imaging equipped on the same machine. The applicability, repeatability (coefficient of variation [CV]), acquisition time, and LS values were compared using the chi-square or Wilcoxon signed-rank tests. In the development group, the correlation between the LS values obtained by the two techniques was assessed with Spearman correlation coefficients and linear regression analysis. In the validation group, the agreement between the estimated and real LS values was evaluated using a Bland-Altman plot.
Results
ElastQ Imaging had higher applicability (94.0% vs. 78.3%, P<0.001) and higher repeatability, with a lower median CV (0.127 vs. 0.164, P<0.001) than did ElastPQ. The median acquisition time of ElastQ Imaging was significantly shorter than that of ElastPQ (45.5 seconds vs. 96.5 seconds, P<0.001). The median LS value obtained using ElastQ Imaging was significantly higher than that obtained using ElastPQ (5.60 kPa vs. 5.23 kPa, P<0.001). The LS values between the two techniques exhibited a strong positive correlation (r=0.851, P<0.001) in the development group. The mean difference and 95% limits of agreement were 0.0 kPa (-3.9 to 3.9 kPa) in the validation group.
Conclusion
ElastQ Imaging may be more reliable and faster than ElastPQ, with strongly correlated LS measurements.
Introduction
Liver fibrosis, which is increasing in prevalence, is one of the most important health issues worldwide [1]. Since liver fibrosis can progress to cirrhosis and lead to hepatocellular carcinoma, it is very important to detect, stage, and monitor liver fibrosis [2,3]. Liver fibrosis is staged using the METAVIR scoring system based on samples obtained by liver biopsy [4]. Although liver biopsy is the gold standard for the assessment of hepatic fibrosis, it is not a suitable tool for monitoring due to its invasiveness and cost. Additionally, sampling errors may occur because of the very small size of the obtained samples and high intra- and inter-observer variability among pathologists [5,6]. Thus, many non-invasive tests, including serum-based markers and elastography, have emerged as alternatives to liver biopsy.
Ultrasonography (US)-based elastography has been regarded as a promising method due to its high diagnostic performance and easy access [7]. This technique is based on the principle that the speed of US waves is faster inside hard tissue than inside soft tissue, making it possible to measure shear wave speed and liver stiffness (LS) values using US-based elastography. Currently, transient elastography (TE) is the most extensively validated form of US-based elastography because it was the first commercially available US-based elastography technique, and many studies have established its high diagnostic performance for liver fibrosis [8-10]. However, TE also has some limitations, including relatively high failure rates in patients with obesity, ascites, and narrow intercostal spaces [11,12].
Unlike TE, newer US-based elastography techniques, including point shear wave elastography (pSWE) and 2-dimensional shear wave elastography (2D-SWE), acquire grayscale images of the liver and can be used to appropriately evaluate LS by placing the region of interest (ROI) using the same probe as that used in the conventional diagnostic US system [13-16]. In addition, pSWE and 2D-SWE have shown comparable or better diagnostic performance and a lower rate of technical failure than TE when evaluating liver fibrosis [1,17-20]. Virtual touch quantification (VTQ; Siemens Healthcare, Erlangen, Germany) of acoustic radiation force impulse was the first pSWE system to be developed and has been validated by many large-scale studies [18,21,22]. A relatively new pSWE technique, ElastPQ, showed good diagnostic performance in predicting significant fibrosis (≥F2) and cirrhosis (F4) with a high success rate and reliable measurements [23-26]. The recent advent of 2D-SWE has allowed for the visualization of multiple shear waves and has enabled both qualitative and quantitative evaluation of LS by providing real-time colored elastographic maps of tissue stiffness. The latest 2D-SWE system, ElastQ Imaging, provides the largest elastographic map available and a unique confidence map that helps physicians select an adequate measuring area. However, few published studies have investigated the clinical applicability of ElastQ Imaging in the evaluation of liver fibrosis.
Therefore, the purpose of this study was to assess the technical performance of ElastQ Imaging compared with ElastPQ equipped on the same US machine, and to investigate the correlation between the LS values obtained using the two techniques.
Materials and Methods
This retrospective study was approved by our institutional review board, and the requirement for informed consent was waived. From March 2017 to December 2018, we searched our radiology database to identify patients who met the following eligibility criteria: (1) being older than 18 years; (2) being referred to the US unit for SWE to determine the presence of liver fibrosis; and (3) undergoing pSWE and 2D-SWE using the same US equipment on the same day. Patients were excluded based on the following criteria: (1) the presence of cofounding factors that might affect LS values (e.g., acute hepatitis, obstructive cholestasis, or hepatic venous congestion) [27]; and (2) insufficient LS measurements (e.g., fewer than 10 measurements for ElastPQ or ElastQ Imaging). A total of 275 patients met all eligibility criteria, and 26 patients were excluded due to the presence of biliary obstruction or heart failure (n=9) and insufficient LS measurements (n=17). As a result, 249 patients were included in our study, and we assessed the technical success rate and reliability of the measurements of the two SWE techniques. After excluding cases with technical failure and unreliable measurements, 189 patients were finally included in our comparison of the repeatability, acquisition time, and LS values obtained by the two SWE techniques (Fig. 1).
The medical records of the enrolled patients were reviewed by two radiologists (M.J.K. and W.H.) with 15 years and 3 years of clinical experience in abdominal imaging, respectively. Clinical and biochemical parameters, including patient age, sex, body mass index, platelet count, and levels of aspartate aminotransferase, were obtained from hospital medical records. The data were obtained within 1 week before or after the LS measurements. The LS values obtained by ElastPQ were used for the assignment of hepatic fibrosis stages, in accordance with the reference values provided by a previous study [24].
LS Measurements Using ElastPQ and ElastQ Imaging
Measurements using both ElastPQ and ElastQ Imaging were made with the same US equipment (EPIQ 7G, Philips Healthcare, Cleveland, OH, USA) by a single operator (S.M.L.) who had 3 years of experience with US elastography and 10 years of experience with US. At least 10 LS measurements were acquired for each SWE technique with the same convex probe (C5-1) used for conventional B-mode US examinations. All patients fasted for longer than 6 hours before the examination and were placed in the supine position with the right arm extended above the head. According to a recently developed guideline, the patients were asked to hold their breath during the LS measurements in order to minimize breathing motions [1]. A measuring box was placed in the right anterior section of the liver, in a location free from areas with artifacts and 1.5-2.0 cm away from Glisson's capsule, using a convex probe and the intercostal approach [1,28].
For LS measurements using ElastPQ, a measuring box of 1.5 cm×0.5 cm was placed in the right anterior section of the liver, avoiding large vessels and bile ducts (Fig. 2A). As ElastQ Imaging was performed, a large, trapezoid-shaped, colored elastographic box was placed in a similar position of the liver with simultaneous use of the confidence map, which shows areas of high confidence in green and areas of low confidence in red. Two or three round ROIs (diameter, 1 cm) were placed within the third elasticity map after passing the first and second elasticity maps (Fig. 2B) [29-31]. To obtain 10 valid measurements, 4-5 breath holds were required when using ElastQ Imaging, whereas more than 10 breath holds are generally required for ElastPQ.
Definition of Applicability, Repeatability, and Acquisition Time
The applicability of the SWE techniques was determined by the ratio of examinations without technical failure or unreliable measurements to all examinations [32,33]. Technical failure of ElastPQ was defined as the inability to obtain 10 valid measurements after at least 15 trials [26]. Technical failure of ElastQ Imaging was defined as the inability to obtain a color map in more than 50% of the sampling area for all acquisitions [34]. For both ElastPQ and ElastQ Imaging, unreliable measurements were defined as those where the ratio of the interquartile range (IQR) to the median of 10 valid LS values was higher than 30% [1,12].
At each session, more than 10 sequential LS values were measured and used to determine the median LS value, the mean LS value, and the standard deviation (SD). The median LS value from each session was regarded as the representative LS value in each patient. To evaluate the repeatability of LS measurements in each session, the coefficient of variation (CV) was calculated as follows: CV=SD/mean LS value [35]. A lower CV indicated higher measurement repeatability.
Acquisition time was defined as the time from the start to the end of obtaining 10 valid LS measurements with each SWE technique. In cases of technical failure (i.e., when 10 valid LS measurements could not be obtained during at least 15 trials), the acquisition time was determined as the time spent measuring all 15 LS values.
Statistical Analysis
Categorical variables (e.g., technical success rate) were compared between the two techniques were compared using the chi-square or Fisher exact test. Results from the Kolmogorov-Smirnov test revealed that all continuous variables (e.g., LS values) were non-normally distributed, and these results were expressed as the median with IQR. The Wilcoxon signed-rank test was used to compare continuous variables between the two techniques. To determine the level of agreement between the LS values obtained using the two techniques, Bland-Altman plot analysis was used [36].
We chronologically divided the enrolled patients into a development group (between March and December 2017) and a validation group (between January and December 2018). Each group was subdivided into two subgroups of patients (with and without significant fibrosis) using a cutoff LS value of 6.8 kPa based on ElastPQ [24]. In the development group and its subgroups, we evaluated the correlation between the LS values obtained by the two SWE techniques using Spearman correlation coefficients and linear regression analysis. A correlation was considered strong if the absolute value of r was >0.7 and moderate if 0.3<r<0.7. We determined the linear regression equation for LS values as ElastQ ( y ) and ElastPQ ( x ). In the validation group and its subgroups, the estimated values of ElastQ Imaging using a linear regression equation were obtained, and we assessed the agreement between the estimated and real LS values obtained by ElastQ Imaging using Bland-Altman plot analysis. All statistical analyses were performed using commercially available software programs (version 24, IBM Corp., Armonk, NY, USA; version 18.11.6, MedCalc Software, Mariakerke, Belgium). For all analyses, P-values of <0.05 were considered to indicate statistical significance.
Results
A total of 249 patients (136 men and 113 women; mean age, 54.4±13.6 years) were included in this study. Clinical history and serological tests for markers such as hepatitis B antigen, hepatitis C antibody, transaminases, alkaline phosphatase, and bilirubin levels revealed evidence of chronic liver disease in 132 (53.0%) of these patients, with the following distribution: chronic hepatitis B (32.5%), chronic alcoholic hepatitis (8.4%), chronic hepatitis C (6.4%), and idiopathic chronic hepatitis (5.6%) [37]. According to the presumed liver fibrosis staging [24], 131 patients (69.3%) had no significant fibrosis (<F2) and the remaining 58 patients (30.7%) had significant fibrosis (≥F2) (Table 1).
Comparison of Applicability between ElastPQ and ElastQ Imaging
Technical failure was not observed in either ElastPQ or ElastQ Imaging in any of the 249 patients. Of the 249 patients, reliable LS measurements were obtained in 78.3% (195 of 249) of the where ElastPQ was used and 94.0% (234 of 249) of the cases where ElastQ Imaging was used (P<0.001).
Comparison of Repeatability, Acquisition Time, and LS Values Obtained Using ElastPQ and ElastQ Imaging
In the 189 patients who had reliable LS measurements in both ElastPQ and ElastQ Imaging, significant differences were found between the two SWE techniques in terms of the CVs, acquisition time, and LS values (Table 2). The CVs obtained using ElastQ Imaging were significantly lower than those obtained using ElastPQ (0.127 [IQR, 0.094 to 0.163] vs. 0.164 [IQR, 0.133 to 0.199], respectively; P<0.001). The median acquisition time of ElastQ Imaging was significantly shorter than that of ElastPQ (45.5 seconds [IQR, 31.0 to 61.0 seconds] vs. 96.5 seconds [IQR, 85.0 to 119.0 seconds], respectively; P<0.001). The median LS value obtained using ElastQ Imaging (5.60 kPa [IQR, 4.66 to 8.34 kPa]) was significantly higher than that obtained using ElastPQ (5.23 kPa [IQR, 4.20 to 7.78 kPa]; P<0.001). The 95% Bland-Altman limit of agreement between the LS values obtained using ElastPQ and ElastQ Imaging was -0.5 of the mean (Fig. 3).
Correlation of LS Values between ElastPQ and ElastQ Imaging in the Development Group
The development group consisted of 146 patients (85 men and 61 women; mean age, 53.1±13.3 years) and was subdivided into subgroups of patients with significant fibrosis (≥6.8 kPa obtained by ElastPQ, n=39) and without significant fibrosis (<6.8 kPa obtained by ElastPQ, n=107). In 146 patients, the LS values for both ElastPQ and ElastQ Imaging exhibited a strong, positive correlation (r=0.851, P<0.001). The linear regression equation for the LS values obtained using ElastQ Imaging ( y ) and those using ElastPQ ( x ) was y =0.86+0.96 x (R2=0.72, P<0.001) (Fig. 4A). Moderate positive correlations between the LS values obtained by the two SWE techniques were observed in the subgroups without significant fibrosis (r=0.494, P<0.001) and with significant fibrosis (r=0.692, P<0.001). In the subgroup without significant fibrosis, the linear regression equation for the LS values obtained using ElastQ Imaging ( y ) and those using ElastPQ ( x ) was y =2.47+0.59 x (R2=0.244, P<0.001) (Fig. 4B). In the subgroup with significant fibrosis, the linear regression equation for the LS values obtained using ElastQ Imaging ( y ) and those using ElastPQ ( x ) was y =1.52+0.92 x (R2=0.48, P<0.001) (Fig. 4C).
Validation of the LS Values Estimated Using the Linear Regression Equation
The validation group consisted of 43 patients (26 men and 17 women; mean age, 58.9±13.2 years) and was subdivided into subgroups of patients with significant fibrosis (≥6.8 kPa obtained by ElastPQ, n=19) and without significant fibrosis (<6.8 kPa obtained by ElastPQ, n=24). The estimated LS values of ElastQ Imaging in the validation group and its subgroups were calculated using the linear regression equation obtained from the development group. The median (IQR) values of the estimated and real LS values for ElastQ Imaging were 5.74 kPa (4.85 to 7.77 kPa) and 5.44 kPa (4.65 to 7.28 kPa), respectively (Table 3). Bland-Altman plots of agreement between the real and estimated LS values are shown in Fig. 5. The mean differences (95% limits of agreement) were 0.0 kPa (-3.9 to 3.9 kPa) in the validation group (Fig. 5A), 0.3 kPa (-2.4 to 2.9 kPa) in the subgroup without significant fibrosis (Fig. 5B), and -1.0 kPa (-12.2 to 10.3 kPa) in the subgroup with significant fibrosis (Fig. 5C).
Discussion
This is the first study to directly compare the applicability, repeatability, acquisition time, and LS values between a 2D-SWE technique (ElastQ Imaging) and a pSWE technique (ElastPQ) using the same US equipment. ElastQ Imaging showed higher applicability, repeatability, and LS values and a faster acquisition time than did ElastPQ. Additionally, the LS values obtained using the two techniques exhibited a strong positive correlation, and the linear regression equation for the LS values obtained using ElastQ Imaging ( y ) and that using ElastPQ ( x ) was y =0.85+0.96 x . However, the estimated values in the validation study were not interchangeable with the real LS values of ElastQ Imaging because of the wide 95% limits of agreement, particularly in the subgroup of patients with significant fibrosis.
In our study, no technical failure was observed for either SWE technique. Nevertheless, the LS measurements obtained using ElastPQ were less reliable than the measurements reported by previous studies [26,38]; however, this discrepancy may have been due to differences in the characteristics of the study populations (e.g., abdominal wall thickness) [39]. Importantly, the LS measurements obtained using ElastQ Imaging were more reliable than those obtained using ElastPQ. Furthermore, ElastQ Imaging showed a lower CV than that of ElastPQ, suggesting higher repeatability. The confidence map of ElastQ Imaging may have contributed to this higher repeatability. Although previous studies have used other maps (e.g., propagation maps) that serve a similar purpose and may improve reliability [31], the confidence map of ElastQ Imaging allows for the visualization of a color-coded value (0%-100%) that indicates the confidence level based on shear wave propagation, while the elastographic map is displayed on the same screen. This allows operators to more easily determine the appropriate location of the ROIs on the elastographic map.
Woo et al. [29] reported that the initially introduced 2D-SWE system, supersonic imaging (SSI), involved a longer measurement duration than did VTQ, the first developed pSWE technique. In contrast, our results demonstrated that the acquisition time of the latest 2D-SWE system, ElastQ Imaging, was significantly shorter than that of the pSWE technique, ElastPQ. The difference in acquisition times could be attributed to the higher B-mode frame rate (approximately 20-30 frames per second) and multiple ROIs in an elastographic box in ElastQ Imaging. Moreover, ElastQ Imaging provides the largest elastographic box among all 2D-SWE systems [29,30], which may enable the reflection of a larger area of liver tissue and more robust fibrosis estimation. Additionally, multiple ROIs could be placed in the box, although studies have yet to determine the precise number of ROIs that could be placed in the elastographic box.
Our results demonstrated that the LS values obtained using ElastQ Imaging were significantly higher than those obtained using ElastPQ, which is similar to the results of previous studies [29,38]. Sporea et al. [38] demonstrated that the mean LS values obtained using ElastPQ were significantly lower than those obtained using VTQ, even though both ElastPQ and VTQ are pSWE techniques. In a study conducted by Woo et al. [29], the LS values obtained using SSI were significantly higher than those obtained using VTQ. Similarly, in a phantom study that compared four different SWE techniques including TE, pSWE, and two types of 2D-SWE, the obtained LS values differed significantly among the techniques [39]. Therefore, distinct cutoff values for discriminating among the different stages of liver fibrosis should be presented for each SWE technique so that they can be used to detect and monitor the disease. However, the cutoff value for ElastQ Imaging for staging hepatic fibrosis using liver biopsy has not been established.
Our study had several limitations. First, this was a retrospective study, and patients with underlying liver disease of various etiologies were included. The causes of liver fibrosis and cirrhosis vary, and this variation may have affected LS values. Second, the diagnostic performance of the two SWE techniques could not be evaluated and compared due to the lack of a reference standard (e.g., a liver biopsy). Further studies of patients with chronic liver disease of a single etiology who undergo liver biopsy as a reference standard are needed to determine the diagnostic performance of the two SWE techniques. Third, we included a large number of patients without chronic liver disease and a small number of patients with significant fibrosis or cirrhosis. Although this reflects actual clinical circumstances, it might have caused spectral bias. In addition, there was a noteworthy difference in the proportion of patients with significant fibrosis (≥F2) between the development group (26.7% [39/146]) and the validation group (44.2% [19/43]). Last, intra- and inter-observer reproducibility was not evaluated, although repeatability was evaluated using CVs.
In conclusion, ElastQ Imaging may be a reliable and fast tool for the evaluation of liver fibrosis. ElastQ Imaging could be a viable alternative to liver biopsy in a clinical setting where rapid and accurate evaluation is required. However, further validation studies using liver biopsy samples as a reference standard are still needed.
Notes
Author Contributions
Conceptualization: Lee SM, Kim MJ. Data acquisition: Lee SM, Kim MJ, Hong W. Data analysis or interpretation: Lee SM, Kim MJ, Yoon JH, Choe JY, Han J. Drafting of the manuscript: Lee SM, Kim MJ, Yoon JH. Critical revision of the manuscript: Ha HI, Lee K, Lee JW, Yoon SY. Approval of the final version of the manuscript: all authors.