Evaluation of the ulnar nerve with shear-wave elastography: a potential sonographic method for the diagnosis of ulnar neuropathy
Article information
Abstract
Purpose
This study was designed to verify whether shear-wave elastography (SWE) can be used to differentiate ulnar neuropathy at the cubital tunnel from asymptomatic ulnar nerve or medial epicondylitis. An additional aim was to determine a cut-off value to identify patients with ulnar neuropathy.
Methods
This study included 10 patients with ulnar neuropathy at the cubital tunnel as confirmed with electromyography (three women and seven men; mean age, 51.9 years), 10 patients with medial epicondylitis (nine women and one man; mean age, 56.1 years), and 37 patients with asymptomatic ulnar nerve and lateral epicondylitis (21 women and 16 men; mean age, 54.0 years). Each patient underwent SWE of the ulnar nerve at the cubital tunnel, distal upper arm, and proximal forearm.
Results
Patients with ulnar neuropathy at the cubital tunnel exhibited significantly greater mean ulnar nerve stiffness at the cubital tunnel (66.8 kPa) than controls with medial epicondylitis (21.2 kPa, P=0.015) or lateral epicondylitis (33.9 kPa, P=0.040). No significant differences were observed between patients and controls with regard to ulnar nerve stiffness at the distal upper arm or the proximal forearm. A stiffness of 31.0 kPa provided 100% specificity, 80.0% sensitivity, 100% positive predictive value, and 83.3% negative predictive value for the differentiation between ulnar neuropathy and medial epicondylitis.
Conclusion
Cubital tunnel syndrome is associated with a stiffer ulnar nerve than lateral or medial epicondylitis. SWE seems to be a new, reliable, and simple quantitative diagnostic technique to aid in the precise diagnosis of ulnar neuropathy at the cubital tunnel.
Introduction
Cubital tunnel syndrome is a common nerve entrapment syndrome of the upper extremity that is caused by compression of the ulnar nerve at the cubital tunnel of the elbow [1]. This condition has a broad symptom complex of sensory and motor deficiencies and dysesthesia [2,3]. Cubital tunnel syndrome is typically diagnosed using the clinical information gained from the patient’s history as well as symptoms, physical examination, and electrodiagnostic studies [4]. However, because cubital tunnel syndrome has a variety of clinical characteristics, it is difficult to differentiate from other diseases that can affect the elbow joint based on clinical presentation and an electrodiagnostic study. Although an electrodiagnostic study is considered the most important diagnostic test for cubital tunnel syndrome, it has some critical disadvantages. It is time-consuming, causes pain and injury to patients due to needle insertion and electrical stimulation, provides limited information regarding lesion localization and structural abnormalities, and carries a strong possibility of falsenegative results [5]. Thus, complementary imaging studies such as magnetic resonance or ultrasound imaging are required for the diagnosis of various neuropathies. Ultrasound imaging in particular supports the diagnosis of neuropathy by providing superior spatial resolution of small peripheral nerves and detailed information regarding lesion localization [6]. Moreover, as a recently-developed ultrasound imaging technology, shear-wave elastography (SWE) has the potential to provide quantitative values for the soft tissue stiffness of tissues, including the muscle [7], tendon [8], joint capsule [9], benign soft tissue mass [10], and nerves, and to improve the diagnostic performance of ultrasound imaging for various neuropathies involving the median, sciatic, tibial, and ulnar nerves [11-15].
The aim of this study was to verify whether cubital tunnel syndrome can be diagnosed based on quantitative ulnar nerve stiffness data obtained with SWE, as well as to compare patients with ulnar neuropathy to those with medial epicondylitis and lateral epicondylitis to establish a precise cut-off value of this parameter for use in ultrasound imaging.
Materials and Methods
Subjects
Institutional review board approval was obtained (CAU IRB No. 1907-013-16575), and the requirement for informed consent was waived because of the retrospective nature of the study. Our study included a review of 77 consecutive patients who had elbow pain and underwent elbow ultrasonography at our institution between October 2018 and December 2018. Twenty patients were excluded due to the lack of an elastography image (n=10), postoperative status (n=2), or a low reliability measurement index (RMI) (RMI≤0.4) (n=8). Therefore, this study included 10 patients with ulnar neuropathy at the cubital tunnel as confirmed with electromyography (three women and seven men; mean age, 51.9±18.8 years), 10 patients with medial epicondylitis (nine women and one man; mean age, 56.1±9.8 years), and 37 patients with asymptomatic ulnar nerve and lateral epicondylitis (21 women and 16 men; mean age, 54.0±8.8 years).
Imaging Protocol
All patients underwent B-mode ultrasound and SWE with a Samsung RS85 ultrasonography apparatus equipped with a 2-to 9-MHz linear transducer (Samsung Medison Co., Ltd., Seoul, Korea). Ultrasound imaging and SWE were performed concurrently by one of two radiologists with more than 10 years of experience in musculoskeletal radiology. The examining radiologist was blinded to the patient’s clinical history and to the results of the physical examination and electrodiagnostic study. The examination was performed with the patient in a sitting position, facing the examiner. The arm to be examined, in the neutral position and with the elbow flexed at 30o, was placed comfortably on a cushion located on the patient’s knee. To minimize the pressure on the skin surface, the probe was coated with sufficient hydrogel and placed on the examined area. During B-mode ultrasound examination of the elbow, the transducer was kept perpendicular to the ulnar nerve. The cross-sectional area (CSA) of the ulnar nerve was measured using a manual tracing measurement tool within the hyperechoic line that surrounds the ulnar nerve. This parameter was measured at three levels of the cubital tunnel: at the medial epicondyle, 4 cm proximal to the epicondyle, and 4 cm distal to the epicondyle. The echogenicity of the ulnar nerve at the cubital tunnel was assessed on a transverse B-mode image for the evaluation of ulnar nerve neuropathy. Ulnar nerve stiffness was measured at three levels along a longitudinal axis of the ulnar nerve: in the middle of the cubital tunnel at the level of the medial epicondyle, 4 cm proximal, and 4 cm distal to the cubital tunnel at the level of the medial epicondyle (i.e., on the distal upper arm and proximal forearm, respectively). The transducer was kept parallel to the skin surface, and special attention was paid to avoid direct compression of the ulnar nerve. In addition, the ulnar nerve was kept in the center of the selected square region of interest (ROI, 5 mm in diameter). Shear modulus data for the selected square ROI were acquired automatically by the ultrasonographic software. The nerve stiffness at each level was measured twice at an interval of 2-3 minutes, and the measurement with the higher RMI value was used as the nerve stiffness value. The results were expressed in kilopascals (kPa). Representative images of the control group, characterized by low values for nerve stiffness, are shown in Fig. 1.
Statistical Analysis
The normality of the distribution of the data associated with the study variables was verified with the Shapiro-Wilk test. The significance of intergroup differences in sex, the affected side, and the echogenicity of the ulnar nerve at the cubital tunnel was analyzed with the Fisher exact test, and that of differences in the numerical characteristics of continuous variables was verified with the Kruskal-Wallis test with the Dunn post-hoc test, the t test, or the Mann-Whitney test. The relationship between CSA and SWE values was evaluated using the Pearson correlation coefficient. The diagnostic accuracy of ulnar nerve stiffness and CSA in the detection of ulnar neuropathy was verified using receiver operating characteristic (ROC) curve analysis. Sensitivity, specificity, and positive and negative predictive values (PPV and NPV, respectively) of each potential predictor of ulnar neuropathy were calculated, along with the area under the ROC curve (AUC) and the associated 95% confidence intervals (CIs). All statistical analysis was performed with statistical software (SPSS version 25.0, IBM Corp., Armonk, NY, USA), with the threshold for statistical significance set at P≤0.05.
Results
The demographic information of the patients is summarized in Table 1. The results of the CSA measurements in the patients with ulnar neuropathy and the controls are presented in Table 2. No statistically significant differences in the CSA of the ulnar nerve at the cubital tunnel were present among the three groups (P=0.081). However, patients with ulnar neuropathy at the cubital tunnel had significantly higher CSA (mean, 13.73 mm2) of the ulnar nerve at the cubital tunnel than the control patients (mean, 8.76 mm2; P=0.002). Statistically significant differences were observed between the patients and the control group with regard to the CSAs at the distal upper arm and proximal forearm. Hypoechogenicity of the ulnar nerve at the cubital tunnel was found in eight patients with ulnar neuropathy, 22 with lateral epicondylitis, and six with medial epicondylitis, and no statistically significant difference with regard to this parameter was present among the three groups (P=0.295).
The results of the SWE measurements in the patients with ulnar neuropathy and the controls are presented in Table 3. Statistically significant differences were observed in the SWE values of the ulnar nerve at the cubital tunnel among the three groups (P=0.048). Patients with ulnar neuropathy at the cubital tunnel had significantly higher SWE values (mean, 66.8 kPa) for the ulnar nerve at the cubital tunnel (Fig. 2) than controls with medial epicondylitis (mean, 21.2 kPa; P=0.015) or lateral epicondylitis (mean, 33.9 kPa; P=0.040). No statistically significant differences were found between patients and controls with regard to ulnar nerve stiffness at the distal upper arm or the proximal forearm. In addition, the ratios of tunnel pressure to proximal pressure and tunnel pressure to distal pressure in the patients with ulnar neuropathy were higher than in the controls (P=0.025 and P=0.022, respectively). The Pearson correlation coefficient between the CSA and SWE values indicated a weak positive relationship (r=0.143), but this relationship was not statistically significant (P=0.290).
An ROC curve illustrating the diagnostic accuracy of SWE (Fig. 3) demonstrated an AUC value of 0.736 (95% CI, 0.560 to 0.912). An ulnar nerve stiffness value of 33.5 kPa provided 70.2% specificity, 80.0% sensitivity, 36.4% PPV, and 94.3% NPV for the differentiation between ulnar neuropathy and other diseases. In addition, an ulnar nerve SWE value of 31.0 kPa provided 100% specificity, 80.0% sensitivity, 100% PPV, and 83.3% NPV for the differentiation between ulnar neuropathy and medial epicondylitis. ROC curves representing the diagnostic accuracy of the SWE ratios of tunnel pressure to proximal pressure and tunnel pressure to distal pressure in ulnar neuropathy exhibited AUC values of 0.728 (95% CI, 0.551 to 0.904) and 0.732 (95% CI, 0.555 to 0.909), respectively (Fig. 3). The cut-off ratios of tunnel pressure to proximal pressure and tunnel pressure to distal pressure with 70.0% sensitivity and 74.5% specificity were 2.1 and 1.7, respectively. In contrast, an ROC curve of the diagnostic accuracy of CSA measurement had an AUC value of 0.713 (95% CI, 0.541 to 0.885). A CSA value of 10.0 mm2 provided 60.0% sensitivity, 68.1% specificity, 30.0% PPV, and 89.2% NPV for the differentiation between ulnar neuropathy and other diseases.
Discussion
In this study, patients with ulnar neuropathy at the cubital tunnel (mean, 13.73 mm2 and 66.8 kPa) presented with greater CSA and SWE values of the ulnar nerve at the cubital tunnel than control participants with medial epicondylitis or lateral epicondylitis. Using ROC analysis, the cut-off range for SWE-determined ulnar neuropathy was found to be 31.0 to 33.5 kPa, which provided optimal accuracy and showed comparable diagnostic performance to the CSA value.
The diagnosis of ulnar neuropathy is based primarily on medical history, physical examination, and an electrodiagnostic study, the last of which is traditionally used for confirmation. However, electrodiagnostic studies have a sensitivity of 37%-86% [16] and are time-consuming and invasive procedures. Therefore, ultrasound imaging may be helpful as an alternative or complementary study in the diagnosis of compressive neuropathy. On B-mode ultrasound images, compressive neuropathy can be diagnosed based on the typical morphological features of compressive neuropathy, including hypoechoic thickening proximal to the compression site, loss of the normal fascicular pattern, and perineural soft tissue edema. Several quantitative ultrasound studies for the diagnosis of neuropathy have been conducted with a focus on CSA measurements [17-19]. These studies have suggested that reference CSA values for the ulnar nerve at the cubital tunnel range from 6 to 9 mm2, and the frequently-used cut-off value is 10 mm2, which has been found to have a broad range of sensitivity ranging from 24.5% to 80% [17]. Our results were similar to the previous findings. In the present study, the cut-off value of CSA was 10.0 mm2, and this cut-off value provided 68.1% specificity and 60.0% sensitivity for the differentiation between ulnar neuropathy and other diseases.
SWE is an advanced quantitative ultrasound technique that can be used to evaluate soft tissue elasticity. In compressive neuropathy, a higher pressure within the canal could cause ischemia, edema, inflammation, and finally fibrosis in the intraneural space and the synovium [20]. All of these changes to the compressed nerve may result in increased nerve stiffness. Therefore, SWE could be a valuable complement in the diagnosis of compressive neuropathy. Some studies have been conducted to analyze the potential application of elastography in the assessment of compressive neuropathies such as carpal tunnel syndrome, ulnar tunnel syndrome, and cubital tunnel syndrome [13,21-24]. Most of these evaluations of compressive neuropathies have been focused on carpal tunnel syndrome. One article demonstrated a clinical application of elastography in cases of cubital tunnel syndrome, with a mean SWE value of 96.38 kPa for patients with cubital tunnel syndrome and a mean value of 33.08 kPa for healthy control participants [13]. In comparison, our study showed that the mean SWE value of patients with cubital tunnel syndrome was 66.8 kPa, while those of controls with medial epicondylitis or lateral epicondylitis were 21.2 kPa and 33.9 kPa, respectively. Our results for the control groups were thus similar to the result of the previous study, while the mean SWE values and diagnostic cut-off SWE value of the cubital tunnel syndrome group in our study were lower than previously found. Our study suggested that SWE values of 33.5 kPa provided 70.2% specificity and 80.0% sensitivity for the differentiation between ulnar neuropathy and other diseases, and this diagnostic performance of SWE was higher than that of CSA on B-mode images. Paluch et al. [13] suggested a cut-off value for ulnar neuropathy of 61 kPa, while Dikici et al. [12] suggested a value for tibial nerve diabetic neuropathy of 51 kPa. Our results may have been affected by the disease severity and duration, the patients’ body mass index, or the small number of patients. Moreover, Cingoz et al. [24] and Dikici et al. [12] have suggested that the severity of neuropathy is correlated with SWE values. Our study included only patients who underwent an initial ultrasound, most of whom could be estimated to have had mild neuropathy. Moreover, the use of an automatically determined ROI setting could have influenced the low cut-off SWE value of ulnar neuropathy in our study because this ROI included the ulnar nerve and some perineural soft tissue. Our cut-off SWE value of the ulnar nerve could represent the upper normal limit of ulnar nerve stiffness. Our control results showed that the mean SWE value of the medial epicondylitis group was slightly lower than that of the lateral epicondylitis group. Although no statistically significant difference was observed between these mean SWE values, and the number of patients with medial epicondylitis is small, this finding could suggest that it is unusual for medial epicondylitis to cause structural changes of the cubital tunnel and compression of the ulnar nerve. In our study, the mean SWE value at the proximal level to the cubital tunnel was lower than that at the cubital tunnel, and this result aligns with previous findings [13]. The ratios of tunnel pressure to proximal pressure and tunnel pressure to distal pressure in patients with ulnar neuropathy were higher than in the controls, and the diagnostic performances of these ratios were similar to those of SWE of the ulnar nerve at the cubital tunnel. In addition, these cut-off values for the ratios of tunnel pressure to proximal pressure and tunnel pressure to distal pressure were similar to the ratios of canal pressure to proximal pressure found in previous studies [13,21]. These ratios may present adjusted values that account for patient-specific and machine-specific factors that impact SWE readings.
In this study, eight patients were excluded due to low RMI values, and these patients had clinical findings of low body mass index, inflammatory changes indicative of calcific tendinitis at the common flexor tendon, and a high signal intensity of the ulnar nerve on magnetic resonance imaging without the symptoms of ulnar neuropathy. In general, relative to strain elastography, SWE is considered to be an easier way to obtain objective and reproducible measurements of direct tissue elasticity because of the lack of need for manual compression. However, this could suggest that the reliability of SWE values is influenced by the nature of the adjacent soft tissue.
This study had some limitations, the most critical of which was its small sample size. Our results showed diagnostic accuracy for cubital tunnel syndrome, which was confirmed with clinical symptoms and electrodiagnostic testing used as a reference standard. However, electrodiagnostic testing has been reported to have low sensitivity. It is possible that we failed to include patients with cubital tunnel syndrome whose findings were negative at the time of electrodiagnostic testing. Additionally, the ROI of the SWE measurement of this ultrasonography equipment was preset for easy and simple manipulation, and our SWE measurements may have included the ulnar nerve and the adjacent soft tissue. Our study did not include an assessment of the interobserver agreement for the SWE measurement of the ulnar nerve. Therefore, longitudinal studies are needed to show the progressive changes in nerve stiffness, especially with regard to treatment.
In conclusion, the ulnar nerve is stiffer in patients with cubital tunnel syndrome than in patients with lateral or medial epicondylitis. In addition, SWE seems to be a new reliable and simple quantitative diagnostic technique to aid in the precise diagnosis of ulnar neuropathy at the cubital tunnel.