Assessing the accuracy and reliability of ultrasonographic three-dimensional parathyroid volume measurement in a patient with secondary hyperparathyroidism: a comparison with the two-dimensional conventional method
Article information
Abstract
Purpose
The purpose of this study was to investigate the accuracy and reliability of the semi-automated ultrasonographic volume measurement tool, virtual organ computer-aided analysis (VOCAL), for measuring the volume of parathyroid glands.
Methods
Volume measurements for 40 parathyroid glands were performed in patients with secondary hyperparathyroidism caused by chronic renal failure. The volume of the parathyroid glands was measured twice by experienced radiologists by two-dimensional (2D) and three-dimensional (3D) methods using conventional sonograms and the VOCAL with 30°angle increments before parathyroidectomy. The specimen volume was also measured postoperatively. Intraclass correlation coefficients (ICCs) and the absolute percentage error were used for estimating the reproducibility and accuracy of the two different methods.
Results
The ICC value between two measurements of the 2D method and the 3D method was 0.956 and 0.999, respectively. The mean absolute percentage error of the 2D method and the 3D VOCAL technique was 29.56% and 5.78%, respectively. For accuracy and reliability, the plots of the 3D method showed a more compact distribution than those of the 2D method on the Bland-Altman graph.
Conclusion
The rotational VOCAL method for measuring the parathyroid gland is more accurate and reliable than the conventional 2D measurement. This VOCAL method could be used as a more reliable follow-up imaging modality in a patient with hyperparathyroidism.
Introduction
In patients with chronic renal failure (CRF), secondary hyperparathyroidism (sHPT) is a common problem. With a decrease in calcium receptors in patients with CRF, the serum level of the intact parathyroid hormone (iPTH) increases, and this is positively correlated with the volume of the parathyroid glands [1,2]. If the volume of one or more parathyroid glands is more than 0.5 cm3 or medical treatment such as active vitamin D or calcimimetics fails, the clinician considers parathyroidectomy [3]. Therefore, clinicians want to know a more exact volume of the enlarged parathyroid glands. In addition, the parathyroid volume could be used as the index of response for medical treatment. Thus far, neck ultrasonography (US) has been performed in patients with CRF for an imaging diagnosis of sHPT and the localization of the parathyroid glands prior to parathyroidectomy.
Recently, a three-dimensional (3D) volume measurement technique using US has emerged, and the virtual organ computer-aided analysis (VOCAL, General Electric Medical Systems, Milwaukee, WI, USA) technology, one of the 3D ultrasonographic semi-automated volume measurement tools to provide accurate and reliable volume calculations, is now commercially available. In the VOCAL technique, after acquiring the 3D volume dataset on US scanning using a 3D probe, a computer-aided analysis program extracts multiple imaging planes while rotating along the designated axis. The number of extracted imaging planes may vary according to the rotational degree selected by the user. Then, the volume is calculated semiautomatically on the basis of a contour drawn by the user on the extracted imaging planes. Thus far, many previous studies have demonstrated the high accuracy and reliability of the VOCAL technique in cases involving fetal organs such as the liver [4,5], lung [6,7], brain [8], bladder [9], and embryo [10], and in phantoms [11,12]. However, thus far, this technique has not been used for measuring the parathyroid gland, and no previous study has compared the virtual volume with the real specimen. The purpose of our study is to investigate the accuracy and reliability of VOCAL for measuring the volume of hyperplastic parathyroid glands in patients with CRF, as compared to the two-dimensional (2D) conventional method.
Materials and Methods
Patients
This prospective study was approved by the Institutional Review Board of Korean University Guro Hospital, and all participants provided written informed consent at the beginning of the study. Between January 2011 and January 2013, a total of 73 patients with sHPT who failed medical treatment underwent US examinations for assessing the parathyroid glands. The volume of enlarged parathyroid glands in patients with sHPT was measured by an experienced neuroradiologist A (1st measurement) using the conventional 2D method. At the same time, the same radiologist A scanned the hypertrophied parathyroid glands by using the 3D method. The scanning dataset was transferred to a personal computer, and the radiologist A created a reconstructed 3D image and measured the parathyroid volume. The 2D and 3D volumes of 203 parathyroid glands in 73 patients were measured, and these volume data were provided to the nephrologist. Ultimately, 15 patients with 40 parathyroid glands who decided to undergo parathyroidectomy were included in this study. The day before the operation, the same neuroradiologist A re-measured (second measurement) the volume of the target parathyroid glands by using the 2D and 3D methods, and provided the volume data to the surgeon. The interval between the first and the second measurements was not more than 2-4 weeks. Immediately after the resection of the parathyroid glands, the specimen volume was measured by another experienced radiologist B.
Image Acquisition
All scans were performed using the 4D8C volume transducer on a LOGIQ 9 ultrasound system (GE Healthcare, Wauwatosa, WI, USA). On the 2D ultrasound, the settings of the machine for the 2D measurements were as follows: frequency, 7.5 MHz; gain, 50; dynamic range, 70; density, high; power, 90; and quality, extreme. The 2D volume of the enlarged parathyroid glands was calculated by an ellipsoid equation as suggested by Zimmermann et al. [13] and Vitti et al. [14] using the craniocaudal (CC), mediolateral (ML), and anteroposterior (AP) diameters. The AP and ML diameters were measured on an axial image. The AP diameter was defined as the longest diameter of the perpendicular line against the skin surface, and the ML diameter was defined as the longest diameter of the nodule parallel to the skin line. The CC diameter was measured on a longitudinal image and was defined as the longest diameter of the nodule parallel to the skin line (Fig. 1A). The parathyroid volume was calculated as follows: π/6×CC×ML×AP. It took about 15 seconds to measure the parathyroid volume by using the 2D method.
The settings of the machine for the 3D measurements were as follows: frequency, 7.5 MHz; gain, 50; dynamic range, 70; density, high; power, 90; and quality, extreme. Each radiologist acquired a single 3D ultrasound dataset and transferred this dataset to the personal computer after checking the quality of the data acquisition by using the 3D display mode. This took approximately 3 minutes. After the Digital Imaging and Communications in Medicine (DICOM) data from the 3D scans were loaded, the axial plane was selected as the reference. The contour of the parathyroid gland on the axial plane was drawn, and two calipers were located on the superior and inferior extremities of the target lesion in order to designate the rotational axis. After the radiologist selected the 30° rotational VOCAL method for measuring the parathyroid volume, the program automatically extracted the six sequential planes perpendicular to the reference axial plane. After the manual drawing of the contour of the parathyroid gland on the six acquired planes, the computer calculated the volume of each parathyroid gland automatically and then displayed a reconstructed 3D image of the parathyroid gland with its volume (Fig. 1B). In all, it took approximately 10 minutes to acquire the US data and calculate the parathyroid gland volume by using the 30° rotational VOCAL technique.
Postoperatively, the specimen volume was measured using the 3D method by radiologist B. The specimen was buried in jelly in order to avoid air and was scanned using the 3D probe (Fig. 1C). The scanning dataset was transferred to a personal computer, and the volume was calculated. The schematic representations of the 2D and 3D measurements and the gold standard of a parathyroid gland are shown in Fig. 2.
Statistical Analysis
All statistical analyses were carried out using SAS ver. 9.2 (SAS Institute Inc., Cary, NC, USA). The intraobserver reliability of the 2D and 3D measurements was evaluated by using the intraclass correlation coefficient (ICC) and Bland-Altman graphical approach. The absolute percentage error of each imaging technique was used for comparing the accuracy of the 2D and 3D techniques. The absolute percentage error was defined as the percentage difference between the ultrasonographically measured parathyroid volume and the specimen volume: 100×(|measured vol.-specimen vol.|)/specimen vol. The Bland-Altman graphical approach was also used for an easy comparison of the accuracy of the 2D and 3D measurements.
Results
In all, 15 patients (4 men and 11 women) with sHPT caused by CRF were enrolled (median age, 52.9 years; mean dialysis time before parathyroidectomy, 12.3 years). The mean serum iPTH levels immediately prior to surgery were 213.0 pg/mL (Table 1). The mean volume of the parathyroid glands on the first and the second measurement conducted using the 2D ultrasound were 1.22±1.66 (mean±SD, cm3) and 1.23±1.40 (mean±SD, cm3), respectively. There was no significant difference between the two measurements (P=0.915) (Table 2). The mean volume of the parathyroid glands on the first and second measurement conducted using the 3D ultrasound were 1.26±1.62 (mean±SD, cm3) and 1.25±1.58 (mean±SD, cm3), respectively. There was no significant difference between the two measurements (P=0.411) (Table 2). The mean specimen volume of the 40 enlarged parathyroid glands measured by the 3D VOCAL method was 1.23±1.57 (mean±SD, cm3).
The ICC value of the 2D measurement was 0.956, indicating a reproducibility of 95.6% (Table 3). The Bland-Altman graph shows that the difference between the two measurements was -0.01±0.63 (mean±SD, cm3), and the limits of agreement (mean±1.96SD, cm3) were 1.23 and -1.25 (Fig. 3A). The ICC value of the 3D measurement was 0.999, indicating a reproducibility of 99.9% (Table 3). The Bland-Altman graph showed that the difference between the two measurements was 0.01±0.08 (mean±SD, cm3), and the limits of agreement (mean±1.96SD, cm3) were 0.15 and -0.17 (Fig. 3B).
The mean absolute percentage difference of the 2D ultrasound measurement of the enlarged parathyroid glands was 29.56±17.37 (mean±SD, %), indicating that the mean measurement error of the 2D method was approximately 30% (Table 3). The Bland-Altman graph for the accuracy of the first 2D measurement showed that the difference between the first 2D measurement and the specimen volume was -0.00±0.41 (mean±SD, cm3), and the limits of agreement (mean±1.96SD, cm3) were 0.81 and -0.80 (Fig. 4A). The difference between the second 2D measurement and the specimen volume was 0.00±0.62 (mean±SD, cm3), and the limits of agreement (mean±1.96SD, cm3) were 1.22 and -1.21 (Fig. 4B). The mean absolute percentage difference of the 3D ultrasound measurement of the enlarged parathyroid glands was 5.78±6.84 (mean±SD, %), indicating that the mean measurement error of the 3D method was approximately 6%. The Bland-Altman graph for the accuracy of the first 3D measurement showed that the difference between the first 3D measurement and the specimen volume was 0.04±0.08 (mean±SD, cm3), and the limits of agreement (mean±1.96SD, cm3) were 0.19 and -0.15 (Fig. 4C). The difference between the second 3D measurement and the specimen volume was 0.02±0.06 (mean±SD, cm3), and the limits of agreement (mean±1.96SD, cm3) were 0.13 and -0.18 (Fig. 4D).
Discussion
In the medical field, accurate and reliable volume measurement of internal organs is very important. Traditionally, volume has been measured using the 2D method and such measurement requires several assumptions. The basic assumption is that the target object has an ellipsoidal shape, but most of the organs have an irregular shape. Therefore, 3D volume measurement was introduced and has been actively used in computed tomography (CT) scans. For example, 3D volume measurements of the liver and the renal cortex provide useful clinical information [15-17]. However, the main drawbacks of 3D measurement using CT scans are radiation hazards and adverse effects from iodine contrast agents. The 3D ultrasonographic volume measurement technique was developed recently, and several studies have shown that 3D ultrasonographic volume measurement is accurate and reliable in many organs [4-12]. Further, because of the easy applicability of US examinations without the use of radiation and contrast media injection, a 3D US volume measurement is more attractive than a CT-based measurement.
There are several 3D US volume measurement techniques. We adopted the 30° rotational VOCAL technique on the basis of the following previous studies. Kot et al. [12] compared the 3D parallel planes technique and freehand technique using three phantoms. Both techniques showed high accuracy and reliability, but the 3D multiplanar technique (3D parallel planes technique) was more accurate (accuracy, 81.5%-83.4% vs. 74.7%-84.2%). Another phantom study compared multiplanar (5-mm interval), VOCAL (30°) and extended imaging VOCAL (5, 10, 15, and 20 planes) methods [11]. All three methods were reliable and accurate (average accuracy, 97.11%). Raine-Fenning et al. [18] showed that rotational VOCAL was superior to the multiplanar method for irregularly shaped organs, but there were no significant differences in measurements among the different rotation angles (VOCAL, 6°, 9°, 15°, and 30°). The relatively high reliability and accuracy of the rotational technique can be explained by the fact that in the multiplanar technique, it is difficult to define the starting and the finishing points for measurement. This problem is overcome in the rotational technique. Further, Peixoto-Filho et al. [9] revealed that the use of a 30° rotation step as compared to a 15° rotation step resulted in a significantly faster evaluation while maintaining excellent reliability in both cases. To summarize, among the several 3D measurement techniques, the rotational VOCAL technique is more accurate and reliable than other 3D measurement techniques such as the 3D parallel planes technique, and the 30o rotational step is time-efficient for measuring the organ volume while maintaining sufficient accuracy and reliability. On the basis of these findings, we selected the 30° rotational VOCAL technique.
In our study, both the 2D and 3D methods showed good reliability, but the 3D ultrasound volume measurement method was more reliable than the 2D method. The ICC of the 2D and 3D measurements was 0.956 and 0.999, respectively. The lower reproducibility of the 2D ultrasound may be explained by the fact that the volume was calculated by using the ellipsoid equation. In this equation, subtle variations in the measurements of each of the three dimensions lead to a large difference in the calculated total volume. Further, in the 2D measurement, it is difficult to obtain the same and exact axial and longitudinal planes of the parathyroid gland on each US examination. However, because the 3D ultrasound volume measurement outlines the parathyroid glands and calculates the volume, it provides more reliable volume data. In practice, when the clinician selects the follow-up imaging modality, he/she considers the two important points of the modality, namely reliability and convenience such as time. Thus, our results have clinical implications, described in the two following points. First, the 3D ultrasound volume measurement could be a better follow-up imaging modality in patients with sHPT with respect to high reliability. Clinically, high reliability is very important in patients with sHPT for the purpose of evaluating the response to medical therapy. Depending on the change in the parathyroid volume, clinicians will adjust the medication dose in order to give the minimum dose to minimize side effects while ensuring response to the treatment. Second, the 2D method may be a sufficient tool for the follow-up of the parathyroid volume with respect to time because it has a relatively high ICC (95.6%) and takes less time to perform. Approximately an additional 10 minutes is required for measuring the volume by using the 3D method.
This study showed that the 3D ultrasound volume measurement is more accurate than the 2D method. The mean absolute percentage difference of the 3D and the 2D ultrasound measurement of the enlarged parathyroid glands is approximately 6% and 30%, respectively. These results suggest that the volume data from the 3D measurement could help determine the optimal time for parathyroidectomy. As mentioned above, if the volume of one or more parathyroid glands exceeds 0.5 cm3 and medical treatment fails, the clinician considers parathyroidectomy [3]. Further, the reconstructed 3D images and the volume information of each hyperplastic parathyroid gland may aid the surgeon to build an image of the target parathyroid gland. Finally, it could provide more accurate clinical information about the response to medical treatment.
Our study has some limitations. First, the number of cases that we examined was small, which prevents us from drawing firmer conclusions from this preliminary result. Second, there is a problem with respect to the gold standard. Several previous in vivo studies revealed the high accuracy of the 3D rotational VOCAL method using the phantoms that have the gold standard of the volume [11,12,18]. Ideally, the gold standard for measuring the volume of the parathyroid gland is the water displacement method by dropping the specimen into the water. However, in reality, we did not adopt this method because the specimen could not be perfectly separated from the fat and the adjacent connective tissues. If the volume was measured by the water displacement method, it would include the volume of the non-parathyroid soft tissues. Instead, we measured the specimen volume after immersing the specimen in jelly, which provides a different tissue contrast. Although we recognize that this is not the ideal gold standard, we accept the accuracy of this technique by using this indirect method.
In conclusion, the rotational VOCAL method for measuring the parathyroid gland is more reliable and accurate than the 2D method, and relatively large-scale studies are required for investigating the value of this new approach.
Notes
No potential conflict of interest relevant to this article was reported.